Failure Analysis Tools

The JetEtch Pro Series

Three chemical decapsulation configurations built on a common JetEtch Pro platform. This page compares the configurations so an appropriate one can be identified for a given application. Process principles, package handling and fixture selection are covered separately on the chemical decapsulation page.

JetEtch Pro chemical decapsulation system, showing the operator control panel, the sealed process chamber and the four-bottle reservoir module.
JetEtch Pro system. Source: JetEtch Pro brochure, p.1.

A common platform

All three configurations share the same etcher unit, reservoir arrangement, etchant chemistry, etch timing, recipe storage and rinse control, as published in their respective brochure specification pages. They differ in the process controls layered on top of that platform.

Shared etcher unit

290 h × 290 w × 419 d mm, 17 kg excluding bottle container and accessories.

Shared chemistry

Fuming nitric, fuming sulfuric and concentrated sulfuric, plus mixed-acid ratios at 6:1, 5:1, 4:1 and 3:1 in both directions.

Shared recipe control

1 – 1,800 second etch time in 1.0-second increments, Pulse and Vortex etch modes, 100 stored user-defined programs.

What differs

Bias selection and the stated minimum etchant temperature. Those differences are set out in the comparison below.

Three configurations

Each configuration has its own page carrying the full specification transcribed from its brochure.

JetEtch Pro chemical decapsulation system, showing the operator control panel, the sealed process chamber and the four-bottle reservoir module.

JetEtch Pro

The core platform configuration. Programmable chemical decapsulation using nitric, sulfuric and mixed acids, with user-defined etch time, temperature, rinse and recipe storage.

  • Temperature from 20 °C, etchant dependent
  • No bias-voltage capability
  • Source: JetEtch Pro brochure and published Nisene comparison table
JetEtch Pro details
JetEtch Pro CuPROTECT system, with the CuPROTECT nameplate on the front panel.

JetEtch Pro CuPROTECT

Adds user-selectable electrical bias to the JetEtch Pro platform, supporting process development for integrated circuits with copper bond wires.

  • Bias selection 0.0 – 20 V in 0.1-V increments
  • Temperature from 20 °C, etchant dependent
  • Source: CuPROTECT insert, p.2
JetEtch Pro CuPROTECT details
JetEtch Pro platform system on which the TotalPROTECT configuration is supplied. The TotalPROTECT insert reuses a CuPROTECT-badged photograph, so a neutral platform photograph is shown here instead.

JetEtch Pro TotalPROTECT

Combines bias selection with a cooling feature that allows sub-ambient cooling of the etching acid, lowering the stated minimum etchant temperature.

  • Bias selection 0.0 – 20 V in 0.1-V increments
  • Temperature from 10 °C, etchant dependent
  • Source: TotalPROTECT insert, p.2
JetEtch Pro TotalPROTECT details

Selection guidance

Configuration selection depends on the device under examination and the structures that must survive the process. Nisene reviews the application before a configuration is recommended.

General chemical decapsulation

Where no electrical bias is required during the etch, the core JetEtch Pro configuration covers the published chemistry, timing and temperature range.

Copper-wire devices

Where an applied electrical bias forms part of the process, CuPROTECT adds user-selectable bias between 0.0 and 20 V.

Lower-temperature process development

Where process development below 20 °C is required, TotalPROTECT states a 10 °C lower limit across nitric, sulfuric and mixed acids.

Application review

Package construction, materials and the examination objective determine the recipe and fixturing.

Comparison

Values transcribed from the specification pages of the three approved brochures and the published Nisene comparison table. “Same” means the value published for that configuration matches the JetEtch Pro column.

JetEtch Pro Series comparison. Sources: JetEtch Pro brochure p.12; CuPROTECT insert p.2; TotalPROTECT insert p.2. Specifications are subject to change without notice.
Specification JetEtch Pro CuPROTECT TotalPROTECT
Etcher unit dimensions 290 h × 290 w × 419 d mm Same Same
Weight 17 kg (38 lb) Same Same
Power 350 W at 95–130 VAC or 210–250 VAC Same Same
Etchant volume 1.0 – 10.0 mL/min Same Same
Mixed-acid ratios 6:1, 5:1, 4:1, 3:1 (both directions) Same Same
Etch time 1 – 1,800 s, 1.0-s increments Same Same
Etch modes Pulse Etch, Vortex Etch Same Same
Heat-up time 0 – 120 s, 1.0-s increments Same Same
Recipe capacity 100 user-defined programs Same Same
Temperature — nitric acid 20 – 90 °C 20 – 90 °C 10 – 90 °C
Temperature — mixed acids 20 – 100 °C 20 – 100 °C 10 – 100 °C
Bias selection No 0.0 – 20 V, 0.1-V increments 0.0 – 20 V, 0.1-V increments
Sub-ambient etchant cooling No No Described in the insert as a cooling feature allowing sub-ambient cooling of the etching acid
Rinse options Nitric acid, sulfuric acid, or no rinse Same Same
Reservoirs 4 bottles standard: 2 acid, 2 waste Same Same
Certifications as printed CE, SEMI S-2-93, SEMI S-2-2000 Same Same
Warranty 12-month parts and labour; 5-year pump assembly Same Same

Process, packages and fixturing

This page compares configurations. The underlying process — how chemical decapsulation works, package handling, defined-area processing, fixture and gasket selection, and process considerations — is covered on the chemical decapsulation page.

JetEtch Pro Series documents

Download the available approved documents directly.

Discuss your JetEtch Pro Series application

Tell us about the package, wire-bond material, required exposure area and analysis objective. Nisene will review the application and advise on system configuration, fixture selection and the appropriate evaluation route.

Nisene Technology Group
Based in Watsonville, California, USA.
417A Salinas Road, Watsonville, CA 95076, USA
sales@nisene.com
service@nisene.com
Schedule a meeting

Send an enquiry

Tell us the system, the package type and what you need to achieve, and our applications team will respond.

Email sales@nisene.com Service and RMA

Nisene Technology Group, 417A Salinas Road, Watsonville, CA 95076, USA